安全第一 - 自检程序步骤
自动或 手动开机
4
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Guenter Nissen
Heiko Riemenschneider
Байду номын сангаас
治疗之前
• 冲洗水路
填充水箱 ,冲洗整个水路
6
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Guenter Nissen
Heiko Riemenschneider
自检 – 维修界面
chapter 3 page 36-38
1. 开机后
软件和定标数据检查
ok
上次消毒记录检查
4. 消毒冲洗 消毒:温度、浓度达到要求
+ 冲洗
2. 治疗前自检
safety system安全报警系统 internal hydraulic进水 blood leak漏血传感器 Heater加热器 Temperature温度传感器
Conductivity电导度传感器 Ultrafiltration超滤部分 SAD空气传感器
自检代码
Training document, no update service.
治疗开始前不可跳过自检。 仅在自检程序完全通过后开始治疗程序才是安全可靠的。
Training document, no update service.
02/11/2019
This document may not be copied, partly or whole, or made be accessible to third parties without Authors:
从硬盘启动程序
LLC 初始化硬件
装载程序 装载定标数据
LLS 初始化硬件
装载程序 装载定标数据
启动 Diabus通讯 检查检测板开关K1位置 0 - 2 - 3
启动与开关位置相应的程序
Therapy - TSM - Software Update 报警信息: RAM/ROM Error 或 CRC calibration Error 如果自检失败
Training document, no update service.
02/11/2019
This document may not be copied, partly or whole, or made be accessible to third parties without Authors:
3
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Guenter Nissen
Heiko Riemenschneider
开机初始化
开机信号发送到SMPS
TLC 主板BIOS程序启动
Self Test SW 7.5x
Service Manual - Chapter 3 维修手册第三章
Training document, no update service.
02/11/2019
This document may not be copied, partly or whole, or made be accessible to third parties without Authors:
Training document, no update service.
02/11/2019
This document may not be copied, partly or whole, or made be accessible to third parties without Authors:
1
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Guenter Nissen
Heiko Riemenschneider
常规
通常每次开机后进入治疗前都必须进行自检。 所有关系到安全的功能都将被测试,确保治疗安全进行。
pressure sensors血路压力
3.治疗 同时传感器不间断的监测 并由独立的监测系统处理器计算报警临界值
Training document, no update service.
02/11/2019
This document may not be copied, partly or whole, or made be accessible to third parties without Authors:
Training document, no update service.
02/11/2019
This document may not be copied, partly or whole, or made be accessible to third parties without Authors:
5
our permission and remains the property of B. Braun Medizintechnologie GmbH with all rights.
Guenter Nissen
Heiko Riemenschneider
自检
• 所有测试步骤都由控制板控制完成 • 监测板通过自己的传感器监测整个过程。 • 监测板测试 D24V 和 B24V 电压继电器各 3 秒。 • 打开 24V 继电器,给控制板进行下一步测试提供电压。 • 任一自检步骤如失败将自动循环监测该步骤。 • 维修界面有自检程序代码,显示自检程序进度。